|
|
|
|
|
|
|
|
|
|
|
|
|
|
Back to " HEALTH PHYSICS INSTRUMENTS "
Detectors
Detector Overview
The kinds of detectors commonly used can be categorized as:
The choice of a particular detector type for an application depends upon the X-ray or gamma energy range of interest and the application’s resolution and efficiency requirements. Additional considerations include count rate performance, the suitability of the detector for timing experiments, and of course, price.
Detector Efficiency
The efficiency of a detector is a measure of how many pulses occur for a given number of gamma rays. Various kinds of efficiency definitions are in common use for gamma ray detectors:
Absolute Efficiency
The ratio of the number of counts produced by the detector to the number of
gamma rays emitted by the source (in all directions).
Intrinsic Efficiency
The ratio of the number of pulses produced by the detector to the number of
gamma rays striking the detector.
Relative Efficiency
Efficiency of one detector relative to another; commonly that of a germanium
detector relative to a 3 in. diameter by 3 in. long NaI crystal, each at 25 cm
from a point source, and specified at 1.33 MeV only.
|
|
Full-Energy Peak (or Photopeak) Efficiency Clearly, to be useful, the detector must be capable of absorbing a large fraction of the gamma ray energy. This is accomplished by using a detector of suitable size, or by choosing a detector material of suitable high Z. An example of a full-energy peak efficiency curve for a germanium detector is shown in Figure 1.1 below.
|
|
Figure 1.1 Efficiency Calibration |
Detector Resolution
Resolution is a measure of the width (full width half max) of a single energy peak at a specific energy, either expressed in absolute keV (as with Germanium Detectors), or as a percentage of the energy at that point (Sodium Iodide Detectors). Better (lower) resolution enables the system to more clearly separate the peaks within a spectrum. Figure 1.2 shows two spectra collected from the same source, one using a sodium iodide detector and one using germanium.
|
|
|
Even though this is a rather simple spectrum, the peaks presented by the sodium iodide detector are overlapping to some degree, while those from the germanium detector are clearly separated. In a complex spectrum, with peaks numbering in the hundreds, the use of a Germanium detector becomes mandatory for analysis
|
|
|
|
|
|
|
|
|
|
Copyright © 2004 Sumhar.com All right reserved |
|||||||